21 CFR Part 820 as it stands
Design history file: what FDA's QMSR now says about design records
Updated
FDA's quality rule now incorporates ISO 13485 by reference, so a device maker's design records are kept under that standard's design and development clause rather than a separate section of Part 820.
Ask about the device data Tell us whether it is design and test records or patient output, and where it is kept; the reply rests on the 19 medtech documents behind this site.
What Part 820 contains now
Part 820 incorporates ISO 13485 by reference in 820.7, and where its clauses conflict with the Federal Food, Drug, and Cosmetic Act or its implementing regulations, those will control. Sections 820.20 to 820.30 are reserved, and its definitions speak of the medical device file.
Design and development
820.10(c) requires manufacturers of class II and class III devices, and of the class I devices it lists (those automated with computer software and those in its table), to comply with the requirements in Design and Development, Clause 7.3 of ISO 13485. Their design records sit under that clause.
Confidential records
820.35(d) lets a manufacturer mark records it deems confidential to aid FDA in deciding what may be disclosed under Part 20. Selling or licensing design records to an AI developer is a separate decision the rule does not make for the company; the register of device data rules quotes each provision.